Image reconstruction in ring array ultrasound tomography by means of
enhancing the Sinogram
a Faculty of
Industrial Technology, Indonesia
b Faculty School of Electrical Engineering and Informatics, Institut Teknologi Bandung, Bandung, Indonesia c
b Faculty School of Electrical Engineering and Informatics, Institut Teknologi Bandung, Bandung, Indonesia c
Abstract
The Ring
Array Ultrasound Tomography is the system which has been used for breast cancer
detection. The system is potentially developed due to unmovable mechanically,
safe and obtainable high resolution image. The study means to reconstruct ring
array ultrasound tomography by reposition and spline-interpolation data into
parallel beam Sinogram. The study was conducted with numerical and experimental
object by 16 and 32 transducer positions to obtain 31×31 pixels of images.
Sinogram was repaired by an RMSD method and reconstructed by the SCFBP method.
The results indicate that the methods of Sinogram enhancement can only improve
the image of 0.02%. The study showed that 16 transducers had produced a blur
image and been dissimilar to the object, but 32 transducers could generate a
good image but it is still worse than parallel beam. © 2013 by IJIR (CESER
Publications).